Home > Products > Heating and Cooling Circulator > -80~200℃ Dynamic Temperature Control System

Intelligence creates the future

Keywords: Rotary Evaporator, Heating and Cooling Circulator, Recirculating Chiller

-80~200℃ Dynamic Temperature Control System

Heater chiller is a laboratory instrument and equipment with dual functions of rapid cooling and heating. The use of fully enclosed pipeline design and efficient plate heat exchangers reduces the demand for thermal fluid and improves the thermal energy utilization rate of the system to achieve rapid temperature control.

  • Introduction

  • Specifications

-80~200℃ Dynamic Temperature Control System-80~200℃ Dynamic Temperature Control System-80~200℃ Dynamic Temperature Control System-80~200℃ Dynamic Temperature Control System-80~200℃ Dynamic Temperature Control System-80~200℃ Dynamic Temperature Control System-80~200℃ Dynamic Temperature Control System-80~200℃ Dynamic Temperature Control SystemGDX-80~200 Degrees Cooling Heating Temp Control System For Pharmaceuticals

Model

GDX-802B

GDX-803B

GDX-806B

GDX-809B

GDX-812B

GDX-818B

Technical data

Temperature range

-80~200℃±0.5℃

Ambient temperature

≤30℃

Ambient humidity

85%

VoltageV

220

380Optional 3P 220V/3P 480V

PhaseP

1

3

FrequencyHz

50/60



PowerKw

4

5.5

13

20

22

36

Heating capacityKw

2

3

6

9

12

18

Cooling capacityw

RT

2200

3650

9000

12750

14500

22000

0℃

1800

31500

7900

9800

10500

18500

-40℃

900

2300

4800

6500

8200

10440

-80℃

240

550

1000

1500

2200

3000

Temperature sensor

PT100

Protection

Protection against phase loss, overvoltage and overcurrent, over temperature surge, fluorine deficiency, overheating, leakage, as well as various alarms such as sensor failure, over temperature surge, fluorine deficiency, etc

Compressor

Type

Closed

Brand

Tecumseh / Danfoss

PowerKw

0.75*2

1.5*2

2.25*2

4.5*2

5*2

7.5*2

Quantity

2

Refrigerant

R404AR23

Condenser

Type

Air cooling / water cooling

Fan type

External rotor axial flow fan

Evaporator

Type

Plate heat exchanger



Tank volumeL

7

7

10

17

35

100

External circulation adapter sizeDN

G1/2

G1/2

G1/2

G3/4

G3/4

G1

Circulating pump

Type

Shield pump



Brand

Shanghai Jingyi



Rated headm

10

10

20

20

15

18

Rated flowL/min

15

15

30

40

65

100

Appearance dimension

WeightKg

200

220

255

335

375

550

Chiller size

Lengthmm

770

710

710

970

970

1400

Widthmm

670

610

610

800

800

970

Heightmm

1180

1250

1250

1250

1350

1580

Control system

Control system

Independently developed microcontroller (optional Siemens PLC) P.I.D. Adaptive temperature control

Programming

5 programs can be written, each with up to 40 steps

Control model

Material temperature and equipment outlet temperature can be switched automatically

Protocol

Modbus RTU protocol RS485 interface (optional RS232 Ethernet)

Control panel

7-inch MCGS high-definition screen (optional Siemens screen)

Data export

Support USB flash disk data export, and export temperature data in TXT format

Packing list

2 sets of adapter valves, one roll of raw meal

2 stainless steel insulated hoses

Operation instructions

All data are measured in the factory, and users will have errors in the process of use